We have compiled information that will help you choose a model, so please refer to it as well.
Although the physical properties vary depending on state of deposition and density, a rough guideline is a dielectric constant of about 2. We are currently conducting field tests with various raw materials.
It is also possible to measure objects with low dielectric constants, which were difficult to measure with microwave Level Transmitters (26GHz series). Once we have collected the test data, we would like to announce it on our website.
Actual results in the field are 5m.
In our in-house offline test, we measured 20m.
It depends on size and amount of foams.
It is difficult to say how much bubbles should be present as it varies depending on physical properties of object to be measured, but from experience we have learned that with foams the size of a soap foam (about 5 cm), the liquid level can be measured by passing through the surface.
As an example of small foams, it is easy to understand if you think of fine foams that occur when sea is rough.
For foams close to this condition, we have a track record of measuring the foam surface but not being able to measure the liquid level.
In such cases, there are cases where problem was solved by using >> guide pipes (breakwater tubes), so please refer to them.
If you have any concerns, we can also conduct an on-site test using a demo machine. Please feel free to contact us. We do not accept changes in electrode dimensions.
We propose a Radar Type River Water Level Transmitter.
Even in situations where there is so much steam that the measurement surface cannot be seen, it can pass through the steam and measure it.
However, if a large amount of water condenses at the base of the microwave Level Transmitter's antenna, it will affect transmission and reception of microwaves.
If there is a large amount of condensation, please purge the inside of antenna with dry air or N2 to blow away the condensation.
If object to be measured is powder or slurry, dust may be mixed with steam and adhere to inside of antenna. In that case, please use optional PTFE coated antenna. This has effect of using condensation to wash away the antenna surface.
However, in this case, condition is that there is always condensation. If it dries during use, it will not be as effective, so please be careful when using it.
>>Example of installation in hot spring tank|Level measurement possible even when visibility is poor due to steam
>>Example of introduction to chicken manure silo|Preventing adhesion and reducing cleaning work
It depends on size and amount of foams.
It is difficult to say how much bubbles should be present as it varies depending on physical properties of object to be measured, but from experience we have learned that with foams the size of a soap foam (about 5 cm), the liquid level can be measured by passing through the surface.
As an example of small foams, it is easy to understand if you think of fine foams that occur when sea is rough.
For foams close to this condition, we have a track record of measuring the foam surface but not being able to measure the liquid level.
In such cases, there are cases where problem was solved by using >> guide pipes (breakwater tubes), so please refer to them.
If you have any concerns, we can also conduct an on-site test using a demo machine. Please feel free to contact us. We do not accept changes in electrode dimensions.
Radar Type River Water Level Transmitter communicate with HART communication module used by another manufacturer?
Set purge pressure to 100kpa and purge amount to 14㎥/h.
We recommend a measurement interval of 30 minutes or more.
If measurements are taken frequently (measurement intervals are short), parts such as motors and wire ropes will deteriorate more quickly.
We recommend that you plan to regularly replace consumables.
We recommend a measurement interval of 30 minutes or more.
If measurements are taken frequently (measurement intervals are short), parts such as motors and wire ropes will deteriorate more quickly.
We recommend that you plan to regularly replace consumables.
Basically it's possible. However, we do not evaluate all equipment, so if you have any concerns, please use a preliminary evaluation such as by renting one out.
Basically it's possible. However, we do not evaluate all equipment, so if you have any concerns, please use a preliminary evaluation such as by renting one out.